Precision Riffle
splitters with controlled sample discharge |


|
Available with 3 mm to 20 mm slots width, specified by the client or selected from recommended standard sizes. Standard sizes conform to the Australian Standard AS 1141.2-1999 & ISO 3082:1998(E)
The splitters are separable. The slots width and
quantity can be adjusted to suit laboratory requirement. Design of the slot protecting
lining is covered by Patent N°2010100147.
Splitters are made from Stainless Steel with
an electro polishing finish. They are suitable for
Food, Pharmaceutical and Geotechnical laboratories.
Splitters feature two dust hoods to minimise dust pollution when operated and a swivelled carrying handle. The swivelled handle provides for extra comfort when shifting the unit around the premises.
The dust hoods are conveniently hinged and allow
for easy inspection of the riffles to check for
blockages. By lifting the dust hood the operator
has clear access to quickly rectify this type of
problem.
|

Precision
Splitter animation
|
Available Configurations:
| Splitters |
16 x 3 |
32 x 3 |
16 x 6 |
12 x 7 |
34 x 6 |
30 x 7 |
16 x 10 |
12 x 13 |
16 x 13 |
16 x 20 |
| Number of Slots |
16 |
32 |
16 |
12 |
34 |
30 |
16 |
12 |
16 |
16 |
| Slot Size, mm |
3 |
3 |
6 |
7 |
6 |
7 |
10 |
13 |
13 |
20 |
| Hopper Capacity, L |
0.23 |
0.6 |
0.55 |
0.46 |
3.7 |
3.75 |
2.72 |
2.72 |
3.7 |
5.8 |
| Each Tray Capacity, L |
0.2 |
0.575 |
0.53 |
0.53 |
2.24 |
2.24 |
1.67 |
1.67 |
2.24 |
3.3 |
| Trays per splitter |
2 |
2 |
2 |
2 |
2 |
2 |
2 |
2 |
2 |
2 |
| Splitter Weight, kG |
1.3 |
4.22 |
2.74 |
2.6 |
8.3 |
8.24 |
6.68 |
6.78 |
8.17 |
10.58 |
| Note: |
Plastic Trays |
SS Trays |
SS Trays |
SS Trays |
SS Trays |
SS Trays |
SS Trays |
SS Trays |
SS Trays |
SS Trays |
The riffle splitter on the left is shown with the dust hoods
open and carrying handle in the upright position.
Below is the splitting ratio test of the 16 x 13 mm Precision Riffle Splitter, conducted at Sibelco Australia Limited, South Australia. The testing personnel filled the hopper and proceeded to weigh the material from 3 splits of progressively less material. The results are as follow:
|
Left tray (g) |
Right tray (g) |
Total (g) |
% Left Tray |
% Right Tray |
Split 1 |
3047 |
3059 |
6106 |
49.90 |
50.10 |
Split 2 |
1527 |
1530 |
3057 |
49.95 |
50.05 |
Split 3 |
764 |
765 |
1529 |
49.97 |
50.03 |
Here is another test done at the same laboratory after using one of our splitters for a while
Dear Igor,
I did a test on one of the splitters you made for us. I chose the one that has done the most work. Below are the results from the testing for your reference. There is a little bit of variation but nothing that concerns us.
Left Side (g) |
Right Side (g) |
3238.97 |
3271.31 |
1628.31 |
1641.54 |
818.66 |
821.45 |
409.16 |
410.86 |
203.9 |
205.54 |
101.39 |
102.73 |
50.2 |
51.11 |
We have been very happy with the splitters. I would certainly be happy to act as a reference if you ever need that.
Regards,
Daniel Maluga
|
Three-Tiered Precision Riffle Splitter |


|
An ME&QC exclusive: this Three-tiered splitter is designed for use in laboratories. Ideal for projects where large quantities or big samples need to be reduced in size quickly.
The splitter incorporates all the best in development of a Single tier Precision Riffle splitter, as described in the section above.
The splitter is made from Stainless Steel with electro-polishing finish. Suitable for Food, Pharmaceutical and Geotechnical laboratories.

The riffle splitter is shown with one
of the front covers lifted.
|

Back cover is lifted for riffles inspection |
|
Laboratory
Riffle Splitters welded. (Slot Width from 6 mm to 12
mm) |


|
Main features of the splitters are:
|
Large
Riffle Splitters. (20 mm to 60 mm slot
width) |



|
Main features of the splitters are:
|
Large
Riffle Splitter animation
|
|
Two-Tiered Splitter |

|
An ME&QC exclusive, this two-tiered splitter is designed for use in the field or in the lab. Ideal for drilling projects where large quantities or big samples need to be reduced in size quickly.
The front dust hood is hinged. That allows for easy inspection of the riffles. It is although convenient in case of riffles are getting blocked. By lifting the hood an operator gets clear access to the riffles. That makes rectifying a problem and cleaning between samples a snap.
Standard configuration of the splitting heads is 14 x 30 mm. The width and quantity of the slots can be specified by a customer. Available slot widths are 20mm, 25mm, 30mm and 35mm.
The riffle splitter is made out of Stainless Steel with electro polishing finish. Suitable for Food, Pharmaceutical, Geotechnical laboratories and field work.
|

|

|
The riffle splitter shown with the front cover lifted.
|
Micro Riffle Splitters
|
 |
Available with 16 slots, 3mm slot width.
The splitters are made from stainless steel with an electro polishing finish. They are suitable for Food, Pharmaceutical, Metallurgical, Geotechnical laboratories.
The splitters feature two dust hoods to minimise dust pollution when operated.
The dust hoods are conveniently hinged and allow for easy inspection of the riffles and to check for blockages. By lifting the dust hood the operator has clear access to quickly rectify this type of problem.
|